8 Maintenance
64
Operating Instructions CEREC AF / AF Connect / AI / AI Connect / AI for Rear
8.1.2

Tablet

CAUTION
The tablet must not be connected during use or when it is within the
patient environment. It must not be charged during this time.
When using the tablet, please ensure that the charge level is sufficient
to complete an entire CEREC application. During the treatment, the
tablet must not be charged in the vicinity of the patient in the treatment
unit.
CAUTION
Using protective sleeve
Please ensure that the tablet is covered with a protective cover during
treatment.
Protective covers help to keep blood, saliva, or dirt away from the
surface of the tablet. Crosstex is one of the suppliers of protective
covers under the name iBarrier (see www.crosstex.com). Protective
sleeves from the manufacturer are included with CEREC AF/AI.
NOTE
If a tablet is used, position it so that there is no risk of damage
occurring, for example on a flat surface.
8.1.3

CEREC Omnicam

CAUTION
If the CEREC Omnicam accidentally falls down, check to make sure
that the camera windows and coated sapphire glass are not
damaged. If the CEREC Omnicam has been damaged, it must no
longer be used on patients.
The CEREC Omnicam must be recalibrated.
NOTE
Do not sterilize the CEREC Omnicam or the camera cable!
NOTE
The mirror sleeves are not autoclave sterilizable!
8.1.3.1

General information

The CEREC Omnicam is a very sensitive optical device and must
therefore be handled with the utmost care. Protect the coated sapphire
glass and the camera windows against scratches and clean them with a
clean lint-free cloth and ethanol (commercially available cleaning
alcohol) whenever a film is noticed during imaging.
Dental health care providers are advised to select the reprocessing
method that aligns with their infection control process. An effective
infection control program is practical, reasonable, and reproducible.
Dentsply Sirona
66 37 487 D3652
D3652.201.01.19.02    10.2018