Operating Instructions CEREC AF / AF Connect / AI / AI Connect / AI for Rear
6. Cautiously fill the container to a filling level of 20mm - 25mm (
- 1 in) under the small elongated hole (A) of the mirror sleeve when
immersed. A funnel can be used to assist with filling the liquid. Do
not spill any disinfectant. If spillage occurs please follow
disinfectant's manufacturer's safety indications for cleaning.
7. Insert the mirror sleeve in the correct position (coated sapphire
glass down).
8. Leave the mirror sleeve at least 12 minutes in the disinfectant.
While exceeding the recommended immersion time may not cause
damage to the mirror sleeve, the mirror sleeve should be removed
as soon as possible after the recommended time.
9. Remove the mirror sleeve from the container.
10. Wipe off the mirror sleeve. Use a cotton gauze moistened with tap
water of potable water quality that meets Federal Clean Water
Standards for this purpose.
11. Dry the mirror sleeve using a clean soft, lint-free cloth.
12. Store the mirror sleeve in such a way that it is protected from
contamination until the next use.
13. Carefully re-attach the mirror sleeve and allow it to lock in place
(see „Refitting the mirror sleeve  [→ 66]").