6 Installation and startup
40
Operating Instructions CEREC AF / AF Connect / AI / AI Connect / AI for Rear
WARNING
Hazard for patient and user
If you do not use freely accessible sockets, there is a risk of injury to
patients and users.
➢ Only use sockets which are freely accessible at all times. This
ensures swift disconnection from the power.
1. Connect the PC to the mains voltage with the power cable.
WARNING
Hazard for patient and user
If you use a camera for the CEREC AI/AI Connect/CEREC AI for
Rear other than the supplied CEREC Omnicam, there is a risk of
electric shock for patient and user.
➢ Only use the supplied CEREC Omnicam for CEREC AI/AI
Connect/CEREC AI for Rear. Do not use a CEREC Omnicam
from a CEREC AC/AC Connect or CEREC AF/AF Connect.
2. Carefully insert the connector of the CEREC Omnicam cable into
the coupling on the shorter of both cables at the camera cradle or
the support arm cable of the CEREC AI/AI Connect/CEREC AI for
Rear, watching out for the guide nose.
3. CAUTION!  Trip/fall hazard! Ensure that the supply cable between
the coupling box and the camera cradle remains fastened at all
times so that the cable cannot pose a trip hazard.
Carefully insert the supply cable connector into the coupling of the
longer of the two cables of the camera cradle (or, for CEREC AI/AI
Connect, into the connection box of the treatment unit to connect to
the supply cable on the camera cradle side), watching out for the
guide nose. Connect the other end of the supply cable to the
coupling box (see image below, connection C).
Dentsply Sirona
66 37 487 D3652
D3652.201.01.19.02    10.2018