• Put the parts together again if they are completely
dry and place them in a dry, sealed container.
Note
Please ensure that the parts are completely dried af-
ter cleaning, otherwise the risk of bacterial growth is
increased.
Sterilisation
• Use cold sterilisation solution in accordance with the
manufacturer's instructions.
Drying
• Place the individual parts on a clean, dry and absor-
bent surface and leave them to dry completely (at
least 4 hours).
Durability of materials
• As with any plastic parts, atomisers and their acces-
sories are affected by a certain amount of wear and
tear when used and hygienically prepared on a fre-
quent basis. Over time, this can lead to a change in
the aerosol, which can have a negative effect on the
efficiency of the therapy. We therefore recommend
that you replace the atomiser and other accessories
after a year.
• The following point should be observed when select-
ing the cleaning product or disinfectant: only use a
mild cleaning product or disinfectant in the concen-
tration and volume prescribed by the manufacturer.
Storage
• Do not store in damp conditions (such as in a bath-
room) and do not transport with any damp items.
• When storing and transporting, protect from pro-
longed direct sunlight.
• The accessories can be stowed securely in the ac-
cessory compartment. Store the device in a dry
place, ideally in the original packaging.

9. Disposal

For environmental reasons, do not dispose of
the device in the household waste.
Please dispose of the device in accordance
with EC Directive – WEEE (Waste Electrical and
Electronic Equipment).
If you have any questions, please contact the local
authorities responsible for waste disposal.

10. Troubleshooting

Problem/
Possible cause/remedy
question
The atom-
1. Too much or too little medication
iser does not
in the atomiser.
produce any
Minimum: 2 ml,
aerosol or
Maximum: 8 ml.
produces too
2. Check nozzle for blockages.
little aerosol.
Clean nozzle if necessary (e.g.
by rinsing out). Then start the
atomiser again.
IMPORTANT: Carefully pierce the
fine holes from the underside of
the nozzle only.
3. Atomiser not held vertically.
4. Unsuitable medication fluid
added for nebulisation (e.g. too
viscous).
The medication fluid should be
prescribed by the doctor.
The output is
Kinked hose, clogged filter, too
too low.
much inhalation solution.
What medica-
Please consult your doctor in
tions are
relation to this matter.
suitable for
As a rule, all medication that is
inhaling?
suitable and approved for device
inhalation, can be inhaled.
There is inha-
This is normal and is necessary for
lation solution
technical reasons. Stop inhalation
residue in the
once the atomiser starts to make a
atomiser.
notably different sound.
What should
1. On infants and children, the
be taken into
mask should cover the mouth
account when
and nose to ensure effective
using the
inhalation.
device with
infants and
children?
2. On children, the mask should
also cover both the nose and
mouth. It is not a good idea to
carry out nebulisation on some-
one who is sleeping, as in this
case not enough of the medica-
tion will reach the lungs.
Note: Inhalation should only be
carried out under the supervision
of an adult and with their assis-
tance and the child should not be
left alone.