1. Specifications
Subject to change without notice.
Specifications
Display
Keypad
Reproducibility (± 1 digit)
Measuring error (± 1 digit)
Temperature
Temperature compensation
Correction of salinity and
atmospheric pressure
Standard calibrations
Dimensions of the measuring instrument and ambient conditions
Operating temperature
Storage temperature
Ambient conditions
Weight
Size
Additional technical data
Energy management
Type of protection
Power supply (batteries)
Connector
Warranty
Warranty
sensION+ DO 6_HACH_0710.indd ANG:3
sensION+ DO 6_HACH_0710.indd ANG:3
Liquid crystal, backlit, with pictograms
Membrane, 5 keys
Saturation
0.0 to 199.9% and 200 to 250% (25°C)
Concentration
0.00 to 19.99 mg/L and 20.0 to 22.0 mg/L (25°C)
Dissolved oxygen
≤ 0.2 of the measured value
Temperature
≤ 0.1 °C (≤ 0.1 °F)
Dissolved oxygen
≤ 0.5 of the measured value
Temperature
≤ 0.2 °C (≤ 0.4 °F)
0.0 to 50.0°C (0 to 122°F)
Automatic, with NTC sensor
Automatic, with manual data entry
At 1 or 2 points. 0% and 100% saturation
0 °C to 50 °C (32 °F to 122 °F)
–15 °C to 65 °C (5 °F to 149 °F)
80% relative humidity (non-condensing)
300 g
186 x 73 x 38 mm
Automatic power off after 5 minutes of inactivity
IP 67
Three 1.5V AA batteries. Autonomy of over 500 hours
MP-5, five-contact multipin
3 years
3
14/9/10 10:51:18