3. When a printer or a computer and Terminal is connected, use the
or
to select
Option
Description
Reduced Select Several or One sample as an output format
Standard Select Several or One sample as an output format. Select several:
Users: The user name appears on the printed report (17
characters). Header:The company name can be added as a
header (40 characters) and appears on the print report. Identify
sensor:The sensor model and the sensor serial number can be
added and appears on the print report.
GLP
Select Several or One sample as an output format. Select several:
Users: The user name appears on the printed report (17
characters). Header:The company name can be added as a
header (40 characters) and appears on the print report. Identify
sensor:The sensor model and the sensor serial number can be
added and appears on the print report.
4. When a computer is connected and ComLabo Easy (refer to for more
information) is selected, use the
Option
Description
Users
The user name appears on the printed report (17
characters).
Identify sensor The sensor model and the sensor serial number can be
added and appears on the print report.
Maintenance
W A R N I N G
Multiple hazards. Do not disassemble the instrument for maintenance. If the
internal components must be cleaned or repaired, contact the manufacturer.
or
to select
C A U T I O N
Personal injury hazard. Only qualified personnel should conduct the tasks
described in this section of the manual.
Clean the instrument
N O T I C E
Never use cleaning agents such as turpentine, acetone or similar products to clean
the instrument including the display and accessories.
Clean the exterior of the instrument with a moist cloth and a mild soap
solution.
Clean the probe
Clean the probe as needed. Refer to
more information about cleaning. Refer to the probe documentation for
information about the probe maintenance.
Replace the magnetic stirrer
If the magnetic stirrer does not start, follow the numbered steps to replace
the magnetic stirrer.
Troubleshooting
on page 15 for
English 13