Hach LZY904.98.00002 Instructions D'utilisation page 5

Masquer les pouces Voir aussi pour LZY904.98.00002:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 24
operates as intended. The manufacturer recommends calibration as local regulations require and
after repairs or comprehensive maintenance work.
3.1 Configure the calibration settings
Select the calibration curve, calibration interval, output behavior during calibration and more.
1. Push menu.
2. Select SENSOR SETUP>TU5x00 sc>CALIBRATION>SETUP.
3. Select an option.
Option
Description
MENU GUIDED
Sets menu-guided calibration to SEALED VIAL, SYRINGE or OFF (default). Calibration
instructions show on the controller display
SYRINGE.
Note: The MENU GUIDED option does not show when sealed vials with RFID are used.
2
CAL CURVE
Selects the type of standard and the calibration curve (range).
STABLCAL 0–40 FNU (default)—1-point calibration (20 FNU) with StablCal.
STABLCAL 0–1000 FNU—2-point calibration (20 FNU and 600 FNU) with StablCal.
FORMAZIN 0–40 FNU—2-point calibration (20 FNU and dilution water) with Formazin.
FORMAZIN 0–1000 FNU—3-point calibration (20 FNU and 600 FNU and dilution water)
with Formazin.
CUSTOM—2- to 6-point calibration (0.02 to 1000 FNU) with StablCal or Formazin. The
user selects the number of calibration points and the value of each calibration point.
STABLCAL 0–40 NTU (or 0–40 FNU) (default)—1-point calibration (20 NTU or 20 FNU)
with StablCal.
STABLCAL 0–700 NTU (or 0–1000 FNU)—2-point calibration (20 NTU and 600 NTU or
20 FNU and 600 FNU) with StablCal.
FORMAZIN 0–40 NTU (or 0–40 FNU)—2-point calibration (20 NTU and dilution water or
20 FNU and dilution water) with Formazin.
FORMAZIN 0–700 NTU (or 0–1000 FNU)—3-point calibration (20 NTU and 600 NTU and
dilution water or 20 FNU and 600 FNU and dilution water) with Formazin.
CUSTOM—2- to 6-point calibration (0.02 to 700 NTU or 0.02 to 1000 FNU) with StablCal
or Formazin. The user selects the number of calibration points and the value of each
calibration point.
VER AFTER
Sets the instrument to start a verification immediately after the instrument is calibrated.
CAL
When set to on, the verification standard is measured immediately after a calibration is
done. Default: ON. Refer to
CAL
Sets the time interval between calibrations. The controller will show a reminder when a
REMINDER
calibration is due. When a calibration is done, the calibration time is set to zero. Options:
OFF(default), 1 day, 7 days, 30 days or 90 days.
OUTPUT MODE Selects the output behavior during calibration. ACTIVE—The outputs continues to give the
measurement values during calibration. HOLD (default)—Keeps the outputs at the last
measurement value before calibration. The outputs give the measurement values again
when the calibration procedure is complete. SET TRANSFER—Sets the outputs to the
SET TRANSFER value selected in the controller settings. Refer to the controller setting for
more information.
CAL POINTS
When the CAL CURVE setting is set to CUSTOM, this option sets the number of
calibration points (2 to 6). This option only shows when the CAL CURVE setting is set to
CUSTOM.
1
Or the Claros user interface for Claros controllers without a display.
2
Select the correct setting for the calibration with StablCal vials with RFID procedure. Refer to
the applicable section of this manual.
1
during calibration when set to SEALED VIAL or
Configure the verification settings
on page 20.
English 5

Publicité

Table des Matières
loading

Table des Matières