Guide and declaration of the manufacturer - electromagnetic immunity
The standard device « Sopix² dental digital x-ray system » is intended to be used in the electromagnetic
environment specified below. The user should make sure it is used in this environment
Immunity trial
Severity level
Conducted RF
3 Vrms
EN 61000-4-6
150 kHz
to 80 MHz
Radiated RF
3 V/m
EN 61000-4-3
80 MHz to 2.5
GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency band applies.
Note 2: These recommendations may not apply in every situation. Electromagnetic wave propagation is
modified by the absorption and reflection due to the structures, objects and persons.
10
²
S o p i x
• U s e r m a n u a l
E N G L I S H
CEI 60601
Compliance
level
3V
3V/m
Electromagnetic environment
Guide
Portable and mobile RF communication devices
should not be used at a distance from the standard
device « Sopix² dental digital x-ray system »,
including the cables, lower than the recommended
separation distance, calculated with the
applicable formulas depending on the emitter
frequency.
Recommended separation distance
d = 1.16 √ P
d = 1.16 √ P 80 MHz to 800 MHz
d = 2.33 √ P 800 MHz to 2.5 GHz
where
is the maximum rated output of the
transmitter in watts (W) by the transmitter
manufacturer and d the recommended separation
distance in metres (m).
The field levels emitted by the fixed RF
transmitters, determined by an electromagnetic
measurement of the sitea, should be lower than
the compliance level in each frequency band.
Interference may occur in the vicinity of the
devices bearing the following symbol: