3 Installation and Commissioning

3
Installation and Commissioning
Operate the medical device exclusively with dental motor SciCan type E-STATIS SLM
and power supply type AEM60US36.

Location

Product installation can be on top, bottom or side of the dental unit in an accessible location.

Connection

Damage as a result of incorrect pressures.
Motor or instrument defect.
Configure pressures in accordance with technical data!
Damage as a result of poor quality media.
Defect of motor or attachment.
ISO 7494-2 states that compressed air must be clean, dry and free from oil.
Note:
If necessary, insert a filter, water trap or air dryer.
Air requirements - see also: chapter 8 Specifications
Measure cooling air quantity at motor coupling
see also: chapter 4 Installation and Commissioning "Measure cooling air quantity at motor coupling"

Connect Power Supply

Note:
The power supply must be connected in compliance with country specific regulations and requirements for
medical devices.
Connect the power supply to the back of the unit.
When switching off the dental unit, the plug of the power supply cord has to be disconnected from the
mains.