Operation; General Operation - SciCan E-STATIS Prep 40-P Manuel De L'opérateur

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4 Operation

4
Operation

General Operation

Parameters configured incorrectly
Damage as a result of incorrect input values.
Check all input values before use. In preparation mode, the motor speed or drill speeds are displayed
according to the speed increase or reduction ratio. In Endodontic mode, the drill speed, torque and drill
direction are shown on the display.
Gearbox ratio incorrect
Damage caused by as a result of incorrect speed/torque.
Select the corresponding transfer factor from the menu and store.
Damage resulting from use of a non authorized transformer.
Damage of product.
Only operate the medical device with the power supply supplied!
Infection Control
After treatment of a patient, let spray air and spray water leak for at least 20 seconds.
Because of stagnation, water- or air-conveying lines in treatment units must be flushed or blown through
before initial operation and after standing times (weekend, public holiday, vacation, etc.).
The sterilization of the medical device must be effected via the treatment unit in the preparation mode.
Switch on medical device
Switch on product by plugging into the electrical circuit. The front circuit board and controller software
versions are displayed on screen for a few seconds. For example:
* software version display
** software version control box
The parameters set (e.g. speed) are then shown on the display.
Note:
Once the foot control has been calibrated, the product is ready for use.
Calibrate foot control - see also: chapter 4 Installation and Commissioning "Calibrate foot control"
Start of the dental motor
Note:
The motor start pressure is 1 bar (14.5 psi).
The minimum operating pressure at 40,000 rpm is 1.8 bar (26 psi).
Press down the foot control pedal until the motor start pressure (1 bar / 14.5 psi) is exceeded.
Motor started.
Press the foot control pedal fully down.
The configured speed is reached.
Page 11
(V xx) OK*
(V xx)
**
E-STATIS Prep 40-P