Calibrate Foot Control - SciCan E-STATIS Prep 40-P Manuel De L'opérateur

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3 Installation and Commissioning

Calibrate Foot Control

Press "Set up" key.
The display then goes in to set up mode.
Pressing the button starts the "psi/bar" calibration process.
Proceed as follows:
1. If "MAX" is not already displayed on the "PSI/BAR" button,
press "PSI/BAR" once or twice until "MAX" appears displayed
on the "PSI/BAR" button.
2. When "MAX" is displayed the unit is ready to activate calibrating
the maximum pressure.
3. Press the foot control ALL THE WAY DOWN and keep holding it down.
Press and HOLD "PSI/BAR" button, (while "MAX" is displayed and foot
control is being held all the way down), until signal sounds.
Maximum pressure has now been calibrated (to equal 100%).
4. The "PSI/BAR" button should now have "MIN" displayed on it.
If it does not, then press "PSI/BAR" once or twice until "MIN" appears
displayed on the "PSI/BAR" button
5. When "MIN" is displayed the unit is ready to activate calibrating
the minimal pressure.
6. DO NOT PRESS THE FOOT CONTROL AT ALL, and
Press and HOLD "PSI/BAR" button, (while "MIN" is displayed),
until signal sounds.
Minimum pressure has now been calibrated (to equal 0%).
Calibration is finished.
Page 10
E-STATIS Prep 40-P
100%
0%