Using the 90DM4
Programming Menus

Main Programming Menu

The MainProg menu provides a number of screens that allow you to
specify the various calibrations, as listed below:
MainProg
Set ADR.
MainProg
Max Conc
MainProg
Out Min
MainProg
Out Max
MainProg
Alarm A
MainProg
Set 4mA
MainProg
Set 20mA
MainProg
Set Zero
MainProg
Set Span
20
Set the unit's communication address.
Adjust the maximum scale concentration for the transmitter
(up to 50000).
Set a minimum concentration for the loop. If set to 750ppm, a
4mA reading is obtained between 0-750ppm. Current
increases proportionally with concentrations above 750ppm.
If you also set an Out Max value, the current increases
proportionally with concentrations between Out Min and Out
Max
Set a maximum concentration for the loop. If set to 1500
ppm, a 20 mA reading is obtained with concentrations
above 1500 ppm . Current increases proportionally with
concentrations between 0 and 1500 ppm . If you also set an
Out Min value, the current increases proportionally with
concentrations between Out Min and Out Max
Alarm setting field.
4mA setting field (adjusts the 4mA loop).
20mA setting field (adjusts the 20mA loop).
Zero calibration field (calibrates the unit's zero).
Span calibration field (calibrates the unit's span).
90DM4 User Manual
Honeywell
511105
2/07