Emissions test
RF CISPR 11 emissions
Harmonic emissions
CEI61000-3-2
Voltage fl uctuation /
Flicker twinkling
CEI61000-3-3
Directives and manufacturer's declaration – Electromagnetic immunity
Powersurge
The
unit is intended to be used in the electromagnetic environment specifi ed below.
The customer or user of the
Powersurge
Immunity test
Test level CEI 60601
Electrostatic
± 6 kV on contact
discharges
± 8 kV in the air
CEI 61000-4-2
± 2 kV for electricity
Rapid transition
supply cables
into salvos
± 1 kV for input/
CEI 61000-4-4
output cables
Transitory excess
± 1 kV between phases
voltage
± 2 kV between phases
and earth
CEI 61000-4-5
<5 % UT
(>95 % cavity of UT)
for 0,5 cycle
Brief cut in voltage
40 % UT
and variations
(60 % cavity of UT)
in voltage in input
for 5 cycles
electricity supply
70 % UT
cables
(30 % cavity of UT)
CEI 61000-4-11
for 25 cycles
<5 % UT
(>95 % cavity of UT)
for 5 s
Magnetic fi eld
for the frequency
of the electricity
3 A/m
network (50/60Hz)
CEI 61000-4-8
NOTE : U
is the alternative voltage before applying the test level
T
Compliance
Class B
The
Class A
including domestic facilities and those directly connected
to the low voltage public electricity network supplying
buildings for domestic use.
Compliance
unit must ensure that it is used in such an environment.
Compliance level
± 6kV on contact
± 8kV in the air
± 2 kV for electricity
supply cables
± 1 kV for input/output
cables
± 1 kV between phases
± 2 kV between phases
and earth
<5 % UT
(>95 % cavity of UT)
for 0,5 cycle
40 % UT
(60 % cavity of UT)
for 5 cycles
70 % UT
(30 % cavity of UT)
for 25 cycles
<5 % UT
(>95 % cavity of UT)
for 5 s
3A/m
BA Powersurge
Premium
Electromagnetic environment - directives
Powersurge
unit is suitable for use in all facilities,
Electromagnetic environment - directives
Floors must be wooden, concrete or ceramic
tiles. If fl oors are covered with synthetic
materials, the relative humidity must be
at least 30 %.
The quality of the electricity supply network
must be that of a typical commercial or hospital
environment.
The quality of the electricity supply network
must be that of a typical commercial or hospital
environment.
The quality of the electricity supply network
must be that of a typical commercial or hospital
environment.
If the user of the
Powersurge
continuous operation during cuts in supply,
it isrecommended supplying the
unit from energy supply with no cut-out or with
a battery.
The magnetic fi elds at the frequency of the
electricity network must have the levels
characteristic of a representative place located
in a typical commercial or hospital environment.
23