7. Adjusting parameters

This is the description of the screen « confi guration » according
Fig.9
to the description
Fig. 9
1
User
Language
Protocol
7
Touch
Pedal
screen

7.1 Language selection

Press the « settings » button
The button
Then select the required fl ag and validate.

7.2 Calibration of contra-angle in use

C
To make the control and display of the torque value more accurate, the contra-angle needs to be calibrated.
This provides the motor with information about output and effi ciency of the contra-angle in use.
The test is carried out in 2 main phases :
w First, the contra-angle is made to rotate with no load.
w The contra-angle must then be connected to the specifi c calibration module. This module simulates an average
intervention in the mouth. The measurement is therefore made in conditions as close as possible to real operation.
w Before using the calibration module, it MUST be lubricated using the
efi tting nozzle.
w Place the adaptor over the connection orifi ce.
w Maintain the spray adaptor in contact on the calibration module.
w Spray for 1 second.
C
The calibration module is not sterilizable.
E
As the capacities of the contra-angles can vary signifi cantly, calibration of the contra-angle is recommended after
maintenance and sterilization, with assistance of a second person to bring the gauge to a standstill (non sterile)
and retrieve the mandrel after calibration.
Press the « settings » button
Press the button
displayed after pressing the button
2
3
4
5
6
Display
Calibration
Light
Sound
function
8
9
10
11
Parameters
Update
Reset
transfer
12
(Fig.1, rep.4 or Fig.3, rep.11).
(Fig.9, rep.2)
accesses the language selection screen.
(Fig.1, rep.4 or Fig.3, rep.11).
Fig.9, rep.3.
.
1. Users or Protocols.
2. Language selection.
3. Calibration of used contra-angle.
4. Adjustment of LED lighting.
5. Adjustment of sound level.
6. Advanced settings.
7. Pedal calibration.
8. Touch pad calibration.
9. Save/ transfer of the parameters.
10. Software update.
11. Re-initialisation of initial confi guration.
12. Return to previous screen.