8. Using the pedal

Button B
Rotation direction
of micromotor
/ Increase in
torque
Button A
Pump
C
If the last protocol sequence is active, a short press on button « C » makes the fi rst sequence of this protocol active.
C
The fi rst protocol sequence is active, a long press on button « C » makes the last sequence of this protocol active.

9. Stopping the device

At the end of the dental surgery :
w Put the supply switch in the O position.
w Remove the tool from the contra-angle.
w Remove the bottle or bag from the bracket.
w Remove the perforator of the irrigation line from the bottle or bag of physiological serum or sterile water.
w Remove the clips for irrigation lines.
w Disconnect the contra-angle from the micromotor.

10. Cleaning, disinfection and sterilisation

10.1 Device maintenance

Single use
w The bags of physiological
fl uid
w The irrigation lines
w Screen protection
w Micromotor protective
sheath
*All of this information is summarised in a maintenance poster supplied with the motor.
Button C
Changing
sequence
Lever
Speed
Non sterilisable
w Control unit
w Pedal
w Calibration module
BA Powersurge
Premium
The lever gradually adjusts the speed according
to the motor during the operational phase.
A short press on the button « A » is a 20% increase in
the speed of the fl ow of the peristaltic pump.
A long press on the button « A » starts the pump at
the programmed fl ow , without using the micromotor,
or the tool.
A short press on the button « B » inverts the rotation
direction of the tool.
A long press on the button « B » increases
the limitation of the torque in increments of 5 N.cm.
A short press on the button « C » goes on to the next
sequence for the active protocol.
A long press on the button « C » goes to the previous
sequence for the active protocol.
Sterilisable in an autoclave
at 134°C for 18 minutes
Micromotor sterilisation cap
Sterilisable
non immersible
(= disinfection)
w Micromotor + cable
Sterilisable
and immersible
(+ ultrasound)
w Bracket
w Motor support
w Mandrel
w Fixing clips for
the irrigation line
19