• STOP - Pressing Enter causes the measurement process to stop immediately
• START - Pressing Enter initiates the measurement process
• STAND-BY - Pressing Enter puts the analyzer in standby
• PROGRAMMING - Program the analyzer (Command mode only)
• READING - View parameters (Operating mode only)
In PROGRAMMING or READING mode the following options can be selected using the scroll keys:
• CALIBRATION - Calibrate the analyzer
• ANALOG-OUT - Analog output parameters
• ALARM - Alarm limit parameters
• SEQUENCE - Measurement mode
• SERVICE - Check the analyzer and accessory operations
• CLEANING - Chemical cleaning parameters
• SAMPLE COND. - Sample conditioning parameters
• CAL PARAMETER - View the primary and last calibration details
• TIMING - Titration cycle parameters
• PM XXXX - System configuration (only available in PROGRAMMING mode and requires a valid
password)
Note: Some of the above options are only available if the relevant accessory has been installed.
7.1 System configuration (PM XXXX option)
Refer also to any Installation Procedure documents delivered with the analyzer for application
specific information.
1. Select PM XXXX with the scrolling keys and press Enter. Using the Select key to move from one
digit to the next, enter the special code and press Enter.
Option
ELECTRODE
UNIT
Adjust ADC
CAL
TEMPERATURE
Adjust Iout
20 English
Description
Define the electrode parameters:
• Name
• Slope
• (Uiso) Isothermal point potential
• (E) Reference point potential adjusted to 25°C
• (C) Reference point concentration adjusted to 25°C
The E and C parameters are used to calculate the standard potential (E
S.logC)
Select the unit of concentration from the list available. Selecting USER allows you to
define your own, and requires a 5 character description.
This option is reserved for Hach Lange service technicians.
Two-point temperature calibration. Use a temperature simulator for each of the two
points. Wait until the measurement stabilizes before accepting each calibration point.
Select the output type (0-20mA or 4-20mA). Place a multimeter in series with the
analog output and make any small adjustments to the signal as required.
= E -