5.4.3.3 Sample level detector
Refer also to
Figure 6
level detector. Wire the connection to J8 as follows:
J8 Terminal number
5.4.3.4 Reagent level detector
Refer also to
Figure 6
level detector. For each reagent, wire the connections to J4, J5, J6 and J7 as follows:
J4, J5, J6 and J7 Terminal number
5.4.3.5 RS232 connection
Refer also to
Figure 6
J10 Terminal number
3 (GND)
The 8810 configuration is:
• Speed: 9600 baud
• Data: 8 bits
• Stop bit: 2
• Parity: none
On startup the analyzer sends the name and software version:
• ANALYZER 8810 : 00 : 00 : 00 :
• ISE X.XX : 00 : 00 : 00 :
In measurement mode, the analyzer sends the following data:
• Measure HH : MM : SS
• M1 XXXXEXX XX.X°C XXX
Where:
• MX = M1 for channel 1 or M2 for channel 2
• XXXXEXX = concentration measurement value
• XX.X°C = temperature
• XXX = value of the potential
on page 14 and
Table 2
1
2
3
4
on page 14 and
Table 2
1
2
3
4
on page 14 and
Table 4
1 (TXD)
2 (RXD
on page 15. The reactor is equipped with a sample
on page 15. Each reagent canister is equipped with a
on page 15. Wire the connection to J10 as follows:
DB9 Plug
RXD: 2
TXD: 3
COM: 5
Color
Brown
Green
Yellow
White
Color
Not used
Not used
Brown
White
DB25 Plug
RXD: 2
TXD: 3
COM: 7
English 17