Technical Information
The minimum rotation speed of the motor is 2200 rpm ±10%* when the control is
between 0-2.6 V. The motor reaches the full speed when the signal is 8.2 to 10 V.
Speed increases proportionally between 2.6 and 8.2 V. See following figure.
Figure 20 RPM. vs. 0-10 V Speed Control Signal
*RPM varies with power supply's voltage. Generally, the higher the power voltage,
the higher the rotating speed. Base pressure decreases and power consumption
increases as the RPM increases.
Grounding Instructions
This product should be grounded. In the event of an electrical short circuit,
grounding reduces the risk of electric shock by providing an escape route for the
electric current. This pump is equipped with a power cord that has a grounding wire
with an appropriate grounding plug. The plug must be inserted into an outlet that is
properly installed and grounded in accordance with all local codes and ordinances.