Instructions for Use
Start-up and Operation of the IDP-3 Dry Scroll Vacuum
Pump
Startup Procedure
1
Do not insert any foreign object through the cooling vents as contact with
WARNING
the fan blades can result in serious personal injury or damage to the pump.
2
Do not block the cooling vents. Blocking these vent holes can cause pump
CAUTION
overheating. A pump surface temperature in excess of 65 °C (150 °F) is
potentially damaging. If such conditions are observed, turn the pump off and
allow it to cool. Disassemble, inspect for damage, and repair, if necessary.
3
The pump ON/OFF switch is a rocker type switch that has symbols in accordance
NOTE
with IEC Publication 417 to represent the ON and OFF positions.
The following figure shows a switch in the ON position.
146
Check that the inlet screen is installed before beginning operation.
Operate the pump at an ambient temperature of 5 °C to 40 °C (41 °F to 104 °F),
otherwise damage to the pump or shortened operating life may result.
Make sure that the pump is configured for the mains voltage to which the
pump is connected.