remarks:
If the blood pressure cuff is not at heart level, a hydrostatic error of about 0.75 mmhg
per centimetre will be produced. If the cuff is lower than heart level, about 0.75 mmhg
per centimetre must be subtracted from the displayed value.
If the cuff is higher than heart level, about 0.75 mmhg per centimetre must be added.
If a repair to this device is necessary, please contact riester directly or the dealer
from whom you purchased the device.
we accept back devices that have been delivered by us and will dispose of them in
a environmentally friendly manner. Please send the device to our company address
(carriage paid by receiver).
the device sticker can display the following symbols:
0124
cemark
do not use in the presence of flammable anaesthetic
patient connection type Bf
serial number
used electrical and electronic devices should not be disposed
of with normal household waste, but should be disposed of
separately according to national or eu guidelines.
device of protection class ii
no alarm function
date of manufacture
oral or axillar measuring point
rectal measuring point
ri-vital
®
warnInGs anD PrecaOutIOns
46
observe the operating manual
sensor not connected to the patient
(displayed in the spo2 window)
sensor not connected to the device
(displayed in the spo2 window)
attach probe cover
not sterile
latex free
single use
dehp free

oVerVieW of functions

4.1 introduction
ri-vital
allows you to measure the patient's oxygen saturation of arterial blood and
®
non-invasive measuring of blood pressure, as well as optional measurement of oral,
axillary and rectal body temperature. the tFt colour display shows the functional
oxygen saturation (spO2) of the arterial blood, the pulse, the mean (MaP), systolic (sYs)
and diastolic (DIa) levels of the arterial blood pressure, as well as the temperature
of the selected region of the body. the sensor used for the spO2 measurement
can be inserted into the front of the device for safekeeping. the optional predictive
thermometer is equipped with a removable/replaceable insertion device for the
thermometer probe, as well as a storage compartment for the probe cover box.
the device has a carrying handle which simultaneously serves as a storage place for
the blood pressure cuffs. the device can be optionally equipped with a holder for the
ri-thermo
n ear thermometer.
®
ri-vital
®
OVerVIew OF FunctIOns
47