Settings
12.5
Battery pack
12.5.1 Battery charge
The gas detector outputs three warning thresholds for the battery charge during
operation.
DIAG_STATUS
Measured
value (read-
ing)
POWER_LOW
available
POWER_LOW,
available
POWER_CRITI-
CAL
POWER_LOW,
NaN
POWER_CRITI-
CAL
12.5.2 Changing the battery pack
For information on changing the battery pack, see "Set up battery operation",
page 39.
13
Settings
13.1
Sensor settings
Sensor settings are available in PolySoft.
50
LED display Time before
Explanation
complete
discharge
none
2-6 months
Order bat-
tery pack.
Yellow LED
Tempera-
Change bat-
flashes
ture depen-
tery pack.
dent approx.
1-3 weeks
(at least 24
operating
hours after
first occur-
rence).
Yellow LED
-
A shut-down
lights up
is possible at
continuously
any time.
Instructions for use
Setting the capture range
The capture range allows unwanted zero deviations to be suppressed. The
setting affects the measured value which is output.
This feature is used to define the capture range and the capture value limits.
Setting the calibration interval
The calibration interval defines the time window for the next calibration. If
the interval expires, the gas detector issues a warning.
Sensor self-test
The sensor self-test is a safety feature that is performed at recurring
intervals. If the sensor does not pass a sensor self-test, a warning is output.
Further negative self-tests will result in an error message being output.
The sensor self-test can be disabled in PolySoft or carried out manually.
For safety-related applications, the sensor self-test must be activated.
Gas settings
Depending on the sensor, the gas type, measuring range and measuring
units can be set with this feature.
13.2
Resetting the gas detector to the factory
settings
Can be performed from 10 minutes after turning on the gas detector.
Device settings, data for ISA100 Wireless
configuration, SIL passwords and Bluetooth
The feature is also available in PolySoft.
Procedure using the magnetic pen
Placement of the magnetic pen
0 s
Reset the gas detector to the factory settings:
1. Hold the magnetic pen on the mark (Fig. A 12) for approx. 15 seconds until
the blue LED flashes rapidly.
2. Lower the magnetic pen for 7 seconds.
®
|
Dräger Polytron
6100 EC WL, Dräger Polytron
network integration, sensor
®
pairing PIN are reset.
LED flashes
15 s
0 s
7 s
8 s
®
Repeater ISA100