Télécharger Imprimer la page

Dräger Polytron 6100 EC WL Notice D'utilisation page 45

Masquer les pouces Voir aussi pour Polytron 6100 EC WL:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 54
Quality
Substatus
Bits 6 and 7
Bits 2 to 4
0
5
0
6
0
7
1
0
1
1
1
2
1
3
1
4
1
5
1
6
2
0
Instructions for use
|
Dräger Polytron
PV
Cause
NaN
No communication with device.
NaN
No communication with device.
NaN
Warm-up phase 1 active.
Maintenance mode active.
Gas reading
– (Bit not used)
Gas reading
– (Bit not used)
Gas reading
– (Bit not used)
Gas reading
– (Bit not used)
Gas reading
Calibration interval expired.
Device is operated outside the specified tempera-
ture range.
Gas reading
Gas reading outside the measuring range.
Gas reading
– (Bit not used)
Gas reading
– (No warning, no error)
®
6100 EC WL, Dräger Polytron
Remedy
A Check ISA100 Wireless
B Check ISA100 Wireless
C Check data for integration into
ISA100 Wireless
A Check ISA100 Wireless
B Check ISA100 Wireless
C Check data for integration into
ISA100 Wireless
Wait for warm-up phase 1 to finish.
Exit maintenance mode.
Calibrate the device.
Use weather protection kit (accessory) when
exposed to sunlight.
A Calibrate the device.
B Change the sensor.
®
Repeater ISA100
|
en
Troubleshooting
receipt.
network settings.
network.
receipt.
network settings.
network.
45

Publicité

loading

Ce manuel est également adapté pour:

Polytron isa100