DIAG_STATUS
21
– (Bit not used)
20
POWER_CRITICAL
19
POWER_LOW
18
SOFTWARE_UPDATE_INCOMPLETE
17
SIMULATION_ACTIVE
16
– (Bit not used)
15
– (Bit not used)
14
OTHER_FAULTS
13
– (Bit not used)
12
– (Bit not used)
11
– (Bit not used)
10
– (Bit not used)
9
– (Bit not used)
8
– (Bit not used)
7
– (Bit not used)
6
SENSOR_WARMUP
5
SENSOR_CAL_INTERVAL_EXPIRED
4
SENSOR_VITALITY_POOR
Instructions for use
|
Dräger Polytron
General status
Cause
F M
Battery charge critical.
M
Battery charge low.
C
Firmware update active.
Firmware update failed.
C
Device is in simulation mode.
F
Serious device fault, various causes.
O
Warm-up phase 1 active.
M
Calibration interval expired.
M
Vitality of the sensor is low.
®
6100 EC WL, Dräger Polytron
Remedy
Change battery pack immediately.
Change battery pack soon.
Wait for firmware update (maximum
15 minutes).
A Restart the device.
B Have the gas detector checked by
DrägerService.
Exit simulation mode via PolySoft.
Contact DrägerService.
Wait for warm-up phase 1 to finish.
Calibrate the sensor.
Change sensor soon.
®
Repeater ISA100
|
en
Troubleshooting
47