Parameter
Available Key Settings
Trigger Source
Key, IO, or Rmt
Trigger Delay
0 to 0.255 s
Repeat Count
1 to 9999
Continuous
-
Step 2 – Run the input sequence
In this example, the load operation is in CV mode.
– Set V
, period, frequency, duty cycle, trigger delay and repeat count accordingly: Use the navigation keys to select
1
the field; use the numeric entry keys to enter the value. The value is set when you press [Enter].
– Set Trigger Source to Key. Press Trig Src Key.
– Press the color-coded [On] key to enable the input.
– Press Run to start the sequence operation. To abort the operation, press Stop.
– Press Back to exit and return to the previous menu.
From the remote interface:
To configure the load operating mode to CV mode:
FUNC VOLT, (@1)
To set the sequencer to Continuous mode:
TRAN:MODE CONT, (@1)
To configure main voltage level with 5 V:
VOLT 5, (@1)
To configure the value for the transient voltage level with 10 V:
VOLT:TLEV 10, (@1)
To configure the transient frequency with 50 Hz :
TRAN:FREQ 50, (@1)
Keysight E36731A User's Guide
Sets the trigger source for the system.
Default is Key.
Action required:
Press Trig Src to toggle between Key, IO, and Rmt.
Key (List Run/Stop Key)
IO (DIO Trigger In)
Rmt (Remote Command)
Sets the trigger delay in seconds.
Default is 0 s.
Sets the list repeat count. This sets the number of times that a list is executed before it
completes.
Default is 1.
Enables check box to repeat the list continuously.
Description
Selects the Run Stopped softkey as a trigger
source.
Selects any configured digital IO (with Trigger
Input function) as a trigger source
Selects a remote interface command as a trigger
source.