Current Limit - You can specify a current limit that limits the input current when in CV mode. It is programmable up
to the rated current, with an over-range of 2%.
Refer to
Constant current mode
Constant resistance mode
In this mode, the load will sink a current linearly proportional to the voltage in accordance with the programmed
resistance value.
Resistance - Lets you enter a resistance value with the numeric keys. Press Enter to enter the value. You can also use
the knob to adjust the value in this field.
Range - Lets you select from three overlapping resistance ranges. The lower ranges provide better resolution at low
resistance settings. Alternatively, you can use the knob or press Enter to toggle between the ranges in this field. See
Programming
Ranges.
Current Limit - The current limit is fixed at the specified setting.
When changing ranges, the load input will be turned off, then back on again.
Refer to
Constant current mode
Constant power mode
In this mode, the load will maintain the input power at the specified programmed power level. Note that the load has
an independent power limit loop that regulates the input power at the limit setting 200 W with an over-range of
10%.
Keysight E36731A User's Guide
for descriptions of the Slew rate, Enable short, Track and Sense controls.
for descriptions of the Slew rate, Enable short, Track and Sense controls.