Specifying the Source/Load Function
You have the option to use the E36731A either as a standalone power supply or electronic load.
When switching between the instrument function, the output is turned off and the Source/Load Settings revert
to their Power-on or RST values.
When switching to Eload mode, there may be a few milivolts residual voltage briefly maintained at the load.
This is due to output capacitor dielectric absorption effect.
1. Press Function to switch the instrument to Load mode or P.Supply mode.
2. You will be prompt with a confirmation message to confirm your selection.
3. Press Yes to confirm your selection. Press No to exit and return to the meter-view display.
From the remote interface:
To switch the instrument to Power Supply mode:
EMUL PSUP
To switch the instrument to Load mode:
EMUL LOAD
70
Keysight E36731A User's Guide