Configuring the Load Settings
Press the Load Settings key to access the Load Settings window.
Operating Modes
Press Mode to select one of the four operating modes (CC, CV, CR and CP). The parameters in Load Settings page
will change accordingly to the selected operating mode. Press Back to exit.
When programmed to a mode, the instrument remains in that mode until the mode is changed or until a fault
condition, such as an overpower or overtemperature, occurs.
Refer to the
Load Operating Modes
When switching between operating modes, the load input is turned off and the Load Settings revert to their
Power-on or RST values.
Constant current mode
In this mode, the load will sink a current in accordance with the programmed value regardless of the input voltage.
Note that a programmable voltage limit is not available. If the DUT imposes a voltage above the voltage rating of the
load, the overvoltage protection will trip.
Current - Lets you enter a current value with the numeric keys. Press Enter to enter the value. You can use the knob
to adjust the value in this field.
Keysight E36731A User's Guide
for more information.