CONTROL PANEL & OPERATIONS
10:00 – Working Time Display. It counts down after work is started.
1.
– Multistage Thermometer. Solid green light indicated that the water temperature
2.
is 0-30% of the set temperature. Solid orange light indicated that the water temperature is 30-
60% of the set temperature. Solid red light indicated that the water temperature is 60-90% of
the set temperature. Flashing red light indicates that the water temperature is close to the set
temperature. All lights indicate that the water temperature is at the set temperature (90-100%).
060 – Set Temperature Display. Press Temp button to select one of the 5 set temperatures
3.
in Celcius or Fahrenheit.
º C º F – ºC or ºF Indicator. ºC - set temperature to Celsius. ºF - set temperature to Fahrenheit.
4.
1:40 – Solution Usage Remaining Time. It counts down to show the remaining usage
5.
time (hour:minute) of the solution.
– Solution Warning Light. Once the Solution Usage Timer is set, the light will come up.
6.
Flashing
, indicates that the solution usage timer is down to zero and the solution needs to be
changed.
– Degas Status. Illuminated, the degas function is selected. Flashing degassing is
7.
underway.
– Normal Cleaning Status.
8.
– Heater Status. Press Heater button twice,
9.
Press Heater button twice again,
Working Status. Green - Ready (G): Normal working status, ready to use. Red - Overheat (R):
10.
Under overheat protection. 20 minutes or more of waiting time is needed for the light to change
from red to green. Work can resume afterwards. If the red light stays on, it indicates possible
malfunction of the transducer. Further checking should be performed.
Heater Key. Press Heater key twice.
11.
heating, press Heater key twice again,
To prevent accidentally turning on or off the heater, the Heater key needs to be touched twice
within 6 seconds to turn it on or off.
When the water reaches the set temperature, the heater will turn off automatically.
Function Key. Press Function key,
12.
On/Off button to start degassing.
normal cleaning function. Pressing Function key before it ends will stop degassing. If additional
degassing is needed after it ends, press Function key again.
Degas: When cleaning heavily soiled or greasy items or silver, copper or brass items, chemical
solutions need to be added into the water. The solution may form many small droplets and take
a long time to dissolve in the water. Newly added water may generate many air bubbles on the
tank walls. These will reduce the cleaning effect in the beginning phase of ultrasonic cleaning.
Turning on the degas function will dissipate the droplets and the air bubbles, usually in 90
seconds, and improve cleaning efficiency.
Illuminated, normal cleaning is underway.
flashes, indicating the heater is turned on.
dims, indicating the heater is turned off.
flashes, indicating the heater is turned on. To cancel
dims, indicating the heater is turned off.
illuminates, indicating Degas function is selected. Press
will flash for 90 seconds, then the unit will return to the