Type
Approved batteries:
Alkaline batteries (ZN-MnO
, U
= 1.5 V)
2
N
Panasonic AA LR6XWA LR6
Duracell Plus Power MN 1500
VARTA 4706 Longlife Max Power AALR6 LR6
Industrial by Duracell AA (LR6) ID1500
Lithium batteries (ZN-MnO
, U
=
2
N
1.5 V)
Energizer Ultimate Lithium AA
Ansmann Industrial Lithium AA
The battery status indicator of the X-act 7000 is designed for alkaline batteries
with a nominal voltage U
of 1.5 V. The use of the specified lithium batteries
N
with a nominal voltage U
of 1.5 V is permissible, but may lead to a false battery
N
status reading.
4.2
Preparation for operation
WARNING
Explosion hazard!
The use of accessories that are not approved for use in explosion-hazard areas
may lead to the ignition of gas-charged atmospheres.
► Only use accessories that have been approved by Dräger for use in
explosion-hazard areas. Further information available on the product page
at www.draeger.com.
Please observe the battery instructions. For further information see:
"Batteries", page 16.
Prior to commissioning, insert the batteries outside of the explosion-hazard area
(Graphic A/13) and attach the carrying strap (see Graphic A/14). Mount the
approved components for probe measurement.
Instructions for use
|
X-act
®
7000
4.2.1
Insert the batteries
1. Unscrew four torx screws on the battery cover (Graphic A/9) on the rear of
the device using the included tool.
2. Insert the batteries included in the scope of delivery (Graphic A/13). Pay
attention to polarity.
3. Close the battery cover again using the included tool. Make sure the battery
cover has been sealed shut.
4.2.2
Attach the carrying strap
1. Attach the carrying strap included to the eye on the rear of the device
(Graphic A/14). Make sure the press studs are on the inside. This prevents
the belt from opening unintentionally.
4.3
During operation
4.3.1
Measuring
Before carrying out the measurement, preparations for the operation must be
completed. For further information see: "Preparation for operation", page 17.
4.3.2
Carry out measurement
1. Turn on the device (Graphic D/3)
● Press and hold OK for 3 seconds until the countdown has run out on the
display.
 The self-test takes place. As soon as this has been successfully
completed, the Insert Dräger MicroTubes icon appears on the display.
2. Insert Dräger MicroTubes (Graphic D/3a)
To do this, place the Dräger MicroTubes in the retainer, as indicated by the
arrow, with the label facing upwards and gently push into place.
|
en
Operation
17