Télécharger Imprimer la page

Dräger X-act 7000 Mode D'emploi page 16

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20
en
|
Operation
3.2
Intended use
The Dräger X-act 7000 is used to determine gaseous compounds in the
ambient air. Depending on the Dräger MicroTubes used, the device measures
the smallest of concentrations of the corresponding substance, predominantly
in the ppb range.
For sensor measurements, a system of sensors and hoses (graphic C/3)
measuring up to 45 meters in length may be connected to the device using the
coupler (graphic C/1) and the X-am pump (graphic C/2).
3.3
Approval
An image of the name plate and the declaration of conformity can be found in
the accompanying supplementary documentation (Part No. 93 00 082).
Intended Operating Area and Operating Conditions, Hazardous Areas
classified by Zones
The device is intended to be used in hazardous areas classified with zone 0,
zone 1 and zone 2, within a temperature range as marked on the device, where
gases of explosion groups IIA, IIB or IIC and temperature class T4 may be
present.
Information about the list of standards to which the device complies is given by
a facsimile reproduction of the EU-Type Certificate and the IECEx Certificate of
Conformity or in the EU-Declaration of Conformity.
If the device is used in off-shore applications, a distance of at least 5 m to any
compasses must be maintained.
16
4
Operation
4.1
Batteries
WARNING
Explosion hazard!
The use of non-approved batteries may cause explosive and combustible
atmospheres to ignite. The intrinsic safety of the device can be endangered.
► Only use the batteries listed in the following chapter, otherwise the approval
will expire.
► Carefully read and follow the description below.
WARNING
Explosion hazard!
The batteries in the device may explode if exposed to intense heat or force.
► Do not throw the device into a fire, keep it away from any sources of heat
and do not try to open it forcefully.
► Only change batteries outside of explosion-hazard areas.
NOTICE
Damage to the device caused by battery acid
Leaking battery acid can damage the device.
► Remove the batteries if the device has not been used for more than 6
months.
► Only insert undamaged batteries into the device.
4.1.1
Requirements
The batteries must meet the following technical requirements:
Type
Explosion protection approval in
accordance with
Temperature class:
Instructions for use
EN/IEC 60079-11
T4 (-20 to +50 °C)
|
®