• Always adjust the lavage so adequate fluid will be
available to cool the tip-tooth interface.
• In general, it is suggested that a "feather light touch" be
used both supra- and subgingivally. The motion of the
activated tip and acoustic effects of the irrigating fluid,
in most cases, is adequate to remove even the most
tenacious calculus.
• Set the System's power adjustment knob to the lowest
power setting for the application and the selected insert.
• Only adjust the System's power adjustment knob with
the insert outside the patients mouth.
10.3 Patient Comfort Considerations
Reasons for sensitivity
• Incorrect tip placement. Point should be directed away
from root surfaces.
• Not keeping the tip in motion on the tooth. Do not allow
the insert to remain in a static position on any one area
of the tooth. Change the insert's path of motion.
• Applying pressure. Use extremely light grasp and
pressure, especially on exposed cementum.
• If sensitivity persists, decrease power setting and/or
move from the sensitive tooth to another and then return.
ENGLISH • 12
Section 11: System Care
11.1 Daily Protocol
Start-up procedures at the beginning of the day
1. Switch ON the dental treatment unit (refer to instruction
manual of treatment unit).
2. Remove the DENTSPLY Cavitron
assembly from handpiece holder and set power
adjustment to 30% of maximum.
®
3.
Connect Steri-Mate
handpiece to the cable connector
assembly (refer to the Infection Control information
booklet for sterilizing instructions).
4.
Adjust the lavage control to maximum.
5.
Hold the handpiece (without an insert installed) over a
sink or drain. Activate the foot control and flush water
through handpiece for at least two minutes.
6.
Place a sterilized insert into the handpiece using a gentle
push-twist motion. Refer to the Cavitron Ultrasonic
Inserts Directions for Use for cleaning and sterilizing
instructions.
7.
Activate the ultrasonics and adjust the power and lavage
control to your preferred operating positions.
Between patients
1.
Remove any ultrasonic inserts and the Steri-Mate
handpiece. Clean and sterilize the handpiece and all
®
cable connector
®