ENGLISH
Operation mode 
Flow rate 
Noise level 
Lamps 
Wavelength 
Ultraviolet energy 
UV‐C radiation external emission 
Equipment 
Reflectors 
Lamp duration 
Germicidal lamp control 
Technical Data of programmer 
Daily and weekly programming of 16 switches (8 x ON, 8 x OFF).  
With day block programming, there are 56 switch possibilities.  
Minimum time resolution between switching points: 1 minute. 
Programs are stored on flash internal memory; the clock is powered from the mains 
and maintained for a few hours by a special battery. 
 
Ambient temperature: ‐10°C to +55°C (T‐10; T55). 
 
STORAGE CONDITIONS
 
‐ Temperature    
‐  Humidity  
‐  Pressure  
 
Before performing installation, make sure the device is intact. Should signs 
of  damage  caused  by  transportation  be  detected,  do  not  activate  the 
 
appliance and inform the dealer. 
 
 
 
 
‐ 10 ÷ 45 °C 
 
10 ÷ 90 % 
 
 500 ÷ 1060 hPa 
 continuous 
3
120 m
/h 
32 dB 
n. 4 UV‐C tubes 25W G13T8 (7 W UVGI) 
253,7 nm 
2
69 µW/cm
 at 1 m (per lamp) 
none 
Dustproof filter; Electronic programmer 
Aluminium 
9000 hours (1 year: 24 hours x 365 dd) 
Anti UV‐C light guide 