ALARM RESPONSE TIMES

Maximum time to alarm (mm:ss)
1 mL/h
Flow rate
Upstream
8:00
Downstream
55:00
Flow rate
1 mL/h
Air
120:00
12

GLOSSARY OF TERMS

BMET
Biomedical Equipment Technician
EMC
Electromagnetic compatibility
IFU
Instructions for Use
IP25
IP Code, International Protection Marking. IP rating determines how resistant something is to environmental
elements. For a medical device, as per the IEC 60529 standard.
Occlusion
A blockage of the system preventing a flow causing a downstream pressure to increase or an upstream pressure
to decrease, activating the pump occlusion out or occlusion in alarm.
PDMS
Patient Data Monitoring System
RF
Radio Frequency
Rotor
Black circular wheel, located on the pump. The Flocare Infinity giving set is attached around the rotor which then
rotates to push fluid or nutrition along the Flocare Infinity giving set.
Occlusion alarm
5 mL/h
50 mL/h
2:30
1:20
11:00
2:00
Air alarm
20 mL/h
100 mL/h
6:00
1:20
- 25 -
100 mL/h
300 mL/h
1:00
0:30
1:00
0:30
600 mL/h
0:30
600 mL/h
0:30
0:30