Alarm description
Air in giving set
Battery empty
Charging please wait
Door is open
Keypad is locked
Last feeding incomplete
Low battery
No plug symbol visible,
while the pump is
connected to the mains.
Occlusion in
Occlusion out
PROBLEM SOLVER CHART
Cause
The pump has detected air in
the giving set.
The air sensor area is dirty
The battery charge is critically
low
The battery is charging.
The door is not properly closed
The keypad is locked to
prevent unintended button
presses
The pump was turned off
before the therapy was
completed.
The battery charge is low
The mains outlet does not
work.
The AC power adapter is
damaged
The pump has detected an
occlusion between the pump
and the feeding reservoir.
The sensor area is dirty
Recurring alarm
The pump has detected an
occlusion between the pump
and the patient.
The sensor area is dirty
Recurring alarm
- 19 -
Solution
If your feeding reservoir is intended to be empty:
Disconnect and discard feeding reservoir and
Flocare Infinity giving set or replace as per
training.
If your feed reservoir is NOT intended to be
empty:
Disconnect from feeding tube and use FILL SET
function to clear air.
Reattach the feeding tube and recommence feed.
Make sure there is no excessive foam in the
feeding reservoir.
Make sure the air sensor is clean.
Connect the pump to the AC power adapter and
mains to recharge the battery immediately for 6
hours.
During charging the pump can be used.
Wait until the battery is charged.
Close the door
Unlock the keypad by pressing i-Button [
SOFT BUTTON 3 simultaneously.
Press START/PAUSE [
]button to start the same
therapy all over.
Press RESUME to continue your active therapy.
Connect the pump to the AC power adapter and
mains to recharge the battery for 6 hours.
During charging the pump can be used.
Connect the pump to another mains outlet.
Contact your pump provider for replacement.
Check the Flocare Infinity giving set between the
pump and the reservoir for kinks or other
obstructions and correct if found. Restart the
pump.
Clean the sensor area
If alarm persists, replace the Flocare Infinity giving
set.
Check the Flocare Infinity giving set between the
pump and feeding tube for kinks or other
obstructions and correct if found.
Check your feeding tube for kinks or other
obstructions according to your training and correct
if found.
Clean the sensor area.
If alarm persists, replace the Flocare Infinity giving
set.