11.3

OPERATING

Atmospheric pressure
Temperature
Humidity
11.4

STORAGE

Atmospheric pressure
Temperature
Humidity
Shelf life
Note: Time to recover from maximum and minimum storage temperature to operating temperature is ~30 min.
11.5

EMC

Electromagnetic emissions may affect the operation of any electronic medical device, including enteral feeding pumps.
The Flocare Infinity III will not be affected by electromagnetic emissions in most environments. However, some
electromagnetic fields produced by personal communication equipment, household appliances, or occupational tools
may cause electromagnetic interference (EMI) which can affect the pump.
Possible sources of electromagnetic interference with electronic medical devices include, but are not limited to:
cellular phones, cordless telephones, microwave ovens, anti-theft/security systems, blenders, and high-powered tools
(i e drills, saws, chain saws).
Check the pump regularly if operating near sources of electromagnetic emissions.
The Flocare Infinity III can safely be operated on commercial aircraft and is designed in accordance with EN
60601-1-2, EN 60601-1-4 for electromagnetic emissions and immunity.
11.6

VOLUMETRIC ACCURACY

Test Condition for measuring volumetric accuracy:
Fluid
Fluid level (head height)
Back Pressure
Temperature
Disposable set
Rate
Volum
Reservoir
Pump position
Feeding Tube
Head height impact
Back pressure impact
65 – 106 kPa
5° - 40°C
15 – 95%, non-condensing
50 – 106 kPa
-20° - 60°C
10 – 95%, non-condensing
None. A normal expect use life is 5 years. The battery life
degrades over time and should be charged before any use.
Tap Water
152 mm ± 76mm, with respect to the centre of the pump rotor
None
Room Temperature [18ºC to 25 ºC]
Flocare Infinity giving set
100 mL/hr
100 mL
Open reservoir
Installed on pole clamp
None
<1.5% per 152 mm
<0.5% per 6.9 kPa
- 24 -