CLEANING, INSPECTION & SERVICING GUIDELINES
10.1

SWITCHING BETWEEN PATIENTS

Enteral feeding pumps must be thoroughly cleaned and inspected before the pump is moved to another patient to
ensure the pump is in pristine condition. We recommend local cleanliness and hygiene policies are followed up.
Clean the pump as described in paragraph 10.2 'Cleaning & disinfection'.
!
Perform inspection as described in paragraph 10.3 'Inspection'
!
Clear the patient history (SETTINGS " RESET PATIENT DATA)
!
If all above actions were taken successfully, the pump can be used by a next user.
10.2
CLEANING & DISINFECTION
For optimum and safe pump operation, the pump should be kept clean and free from debris. It is recommended to
clean the pump with a wet cloth after every completed feeding regimen.
Disconnect the pump from the mains before cleaning. To avoid electrical shock, never clean the
pump with AC power adapter plugged into an outlet or when pump is on.
Do not immerse the pump in water.
Do not use aggressive cleaners or disinfectants such as acetone, benzene or iodine.
Note: the pump is not intended to be cleaned using a dishwasher.
Do not sterilize the pump, e.g. by irradiation, EtO, steam or in an autoclave.
Pump may be cleaned by rinsing with warm soapy water (standard washing up liquid) and a non-abrasive sponge,
brush or soft cloth. Clean and remove any nutritional residue from the pump surface paying particular attention to the
sensor area on the top housing, the rotor including its rollers and in the corners of the keypad and screen where it
meets the pump housing. Use a cotton swab to clean pathways of cassette insert and to remove any silicon residue
from pump wheel roller pins. Then dry the pump with a clean cloth.
The pump has been untouched
for 3 minutes or more.
No rate has been programmed
when starting a therapy.
Giving set not properly
installed
The giving set has been used
for more than 24 hours
Preventive maintenance
service interval has elapsed
Reminder that the giving set
has been used for more than
24 hours and needs
replacement.
The pump has encountered an
internal system failure.
The pump has completed the
delivery of the programmed
therapy.
The therapy has reached its
programmed target volume
No volume has been
programmed when starting a
therapy.
- 20 -
Press any button to acknowledge the warning
message.
Program a Rate (mL/h)
Insert a Flocare Infinity giving set
Replace the Flocare Infinity giving set with a new
one.
Return pump to pump provider for service.
Replace the Flocare Infinity giving set with a new
one.
Turn the pump OFF, make sure the pump door is
closed and switch the pump back ON.
If the error persists, contact pump provider for
replacement.
Press HOME [
] button if your therapy is
completed.
Press the HOME [
] button to acknowledge the
warning message. If your therapy is completed,
turn the pump off.
Program a Volume (mL)