6.4

SET THE LANGUAGE

Set the language by pressing the SETTINGS
Note: at first pump start up or after a factory reset, the pump will guide you automatically through the language
selection.
6.5

SET THE TIME AND DATE

Set the date and time by pressing the SETTINGS
time zone. The pump will not change to daylights savings automatically nor will it change to other time zones during
travelling.
Note: at first pump start up or after a factory reset, the pump will guide you automatically through the date and time
setting.
6.6

SWITCH THE PUMP OFF

Turn the pump OFF by pressing and holding the ON/OFF button for approximately 2 seconds until you hear the turn-
off sound.
7

OPERATING INSTRUCTIONS

7.1

LOAD THE FLOCARE INFINITY GIVING SET

Follow the instructions for use on the packaging of the Flocare Infinity giving set, for setting up the enteral system.
1.
Open the door
2.
Position the purple looped section of the cassette (R) around the pump rotor (D)
3.
Gently stretch (do not over stretch) and seat the cassette into the pump
The cassette is designed to fit into the pump one way
4.
Close the pump door
5.
Pump is now ready to be programmed
7.2

PAUSE / RUNNING MODE

The Flocare Infinity III is either in pause mode waiting for user instructions or it is in running mode delivering a
therapy to a patient. The pause or running modes are controlled by the START/PAUSE [ ] button.
Note: when the pump has been untouched for 3 minutes or more in PAUSE mode, the pump will display the 'Pump
needs attention' alarm message.
7.3

FILL THE FLOCARE INFINITY GIVING SET

The FILL SET function is used to prime a (new) Flocare Infinity giving set with nutrition or water and needs to be
commenced before starting a feeding therapy.
Ensure the Flocare Infinity giving set is disconnected from the patient feeding tube. Prime the set by press and hold
the FILL SET button for 2 seconds to activate the FILL SET function and release the FILL SET button as soon as the
pump generates a beep and automatically starts.
The FILL SET can be manually stopped by pressing STOP in case the giving set is over filled.
Note: when the FILL SET function is activated, all alarms are temporary deactivated helping avoid alarms being
inadvertently triggered (e.g. air alarm).
button, n avigate to LANGUAGE and select your preferred language.