7.7.2

Portions information

When an INTERMITTENT therapy has been programmed, the Flocare Infinity III provides useful information
concerning your individual portion when the PORTIONS INFORMATION button is pressed.
This function can be access from the HOME SCREEN " SOFT BUTTON 1 " PORTIONS INFORMATION
7.7.3

30-day history graph

The Flocare Infinity III stores all administered therapies for a maximum period of 30 days.
The daily administered volume can be accessed by pressing SOFT BUTTON 1 from the home screen and pressing
HISTORY. With the arrow buttons, you can navigate to earlier or later days. By pressing the + magnifying symbol
you can zoom in on your selected day by showing a table of individual therapies.
7.7.4

Reset progress bar

The Progress bar is automatically cleared after completion of each therapy where a VOLUME was programmed.
When there is a need to manually clear the Progress bar, navigate from the HOME screen to SOFT BUTTON 1 "
RESET PROGRESS BAR.
7.8

KEYPAD LOCK

The keypad can be locked to avoid unintentional changes. The keypad lock can be activated by pressing the
INFORMATION button and SOFT BUTTON 3 simultaneously when the pump is in running mode. Follow the same
procedure to unlock the keypad. Icons will guide the user on how to unlock the keypad to support untrained users.
Note: the pump can still be turned OFF even when the keypad is locked.
7.9

CONNECT TO A NURSE CALL OR PDMS SYSTEM

The power connector at the left-hand side of the pump offers the option of connecting the pump to an external alarm
system (Nurse call) or Patient Data Monitoring System (PDMS). For these functions, we offer the following
accessories:
The Flocare® Infinity™ Nurse Call enables connection to an external alarm system. Follow the instructions
!
for use of the Flocare® Infinity™ Nurse Call to connect the pump.
The Flocare® Infinity™ PDMS Cable enables connection to an external PDMS system. Follow the
!
instructions for use of the Flocare® Infinity™ PDMS Cable to connect the pump to an external PDMS
system
7.10

BATTERY

The Flocare Infinity III is equipped with Lithium/Ion battery technology and needs to be fully charged for a period of 6
hours prior to first use, this will maximise the battery life time.
The battery microprocessor continuously monitors the battery state of health and the pump will warn the user when
the battery is depleted for 70% or more and needs replacement.
Battery age
1 year
2 year
3 year
4 year
5 year
70% depleted
Note: The battery capacity is an estimate. If you are unsure that enough battery capacity remains for your intended
therapy, connect to the AC power adapter until recharged.
125 mL/h
24 hrs
24 hrs
20 hrs
16 hrs
12 hrs
7.2 hrs