5.4.3

Progress bar

The progress bar provides a quick overview about the amount of delivered tube feed since the last time the
progress bar was cleared.
Note: refer to chapter
VOLUME - RATE
When a volume - rate therapy has been programmed, the progress bar is fully outlined and gives real-time
information of the administered volume on the left side.
RATE
When a rate therapy has been programmed, the progress bar fades out at the end of the graph indicating
that no volume has been programmed. The graph is static and shows only the administered volume since the
last time it was cleared.
INTERMITTENT
When an intermittent therapy has been programmed, the progress bar is separated into the number of
programmed portions. The progress bar is capable to show up to 3 portions, if more portions are
programmed, they will appear automatically after the previous administered portions has been delivered.
Progress of each individual portion is indicated by filling in real-time the programmed portion comparable to
volume/rate.
5.4.4

Running indicator

The running indicator symbol becomes active after pressing the START/PAUSE [ ]button. The indicator shows 6
dots moving from the left to right and represents an active programmed therapy.
In PAUSE mode, only 3 static dots are visible.
6

PUMP SETUP

6.1

ATTACHING TO THE POLE CLAMP (OPTIONAL)

1.
Connect the Flocare Infinity III pole clamp (N) to a pole stand by rotating the black star shaped bolt clockwise.
2.
Connect the Flocare Infinity III to the pole clamp interface (M) by rotating the grey wheel shaped disk clockwise
until the pole clamp is fully connected to the pump without any clearance.
6.2

CONNECT TO THE AC POWER PLUG

Connect the AC power adapter by gently pressing the charger plug (E) into the pump (J).
The charger plug is designed to fit in only one way. When the pump is connected to the mains, the blue AC charging
light (K) should be on and the plug icon to be appeared next to the battery icon.
Note
:
The Flocare Infinity III is equipped with Lithium/Ion battery technology and needs to be fully charged for a
period of 6 hours prior to first use.
6.3

SWITCH THE PUMP ON

Turn the pump ON by pressing and holding the ON/OFF button for approximately 2 seconds until the start-up sound
can be heard.
The pump will perform a self-test and inform the user when the device is ready to use, the home screen will be shown
after the self-test.
Note: When a previous therapy has not been administered completely before the pump was turned OFF, the pump
will show the warning message 'Last feeding incomplete'. Refer to chapter
for more information.
therapy'
7.7.4 'Reset progress bar'
for a manual reset of the progress bar.
7.5.4 'Change the parameters of an active
- 11 -