Fig. 4
value of 3.87KΩ.
T h e
a c t u a l
value of resis-
tance is 3.87KΩ
and not 4KΩ
that was mea-
s u r e d i n t h e
200KΩ range.
This is very im-
p o r t a n t b e -
c a u s e i f t h e
manufacturer
specifications
s a y t h a t t h e
sensor should
read 3.8-3.9KΩ
at 70°F then on
2 0 0 K Ω
t h e
range the sensor would be defective, but
at the 20KΩ range it would test good.
Now set the multimeter to the 2KΩ range.
(See Fig. 6) The
display will indi-
c a t e
a n
overrange con-
dition because
3.87KΩ is larger
than 2KΩ.
This example
shows that by
decreasing the
range you in-
crease the ac-
curacy of your
measurement.
W h e n
y o u
c h a n g e t h e
r a n g e , y o u
change the lo-
cation of the decimal point. This changes
Fig. 5
Fig. 6
the accuracy of the measurement by ei-
ther increasing or decreasing the number
of digits after the decimal point.
Battery and Fuse
Replacement
Important: A 9 Volt battery must be in-
s t a l l e d b e f o r e u s i n g t h e d i g i t a l
multimeter. (see procedure below for
installation)
Battery Replacement
1. Turn multimeter OFF.
2. Remove test leads from
multimeter.
3. Remove screw from battery
cover.
4. Remove battery cover.
5. Install a new 9 Volt battery.
6. Re-assemble multimeter.
Fuse Replacement
1. Turn multimeter OFF.
2. Remove test leads from
multimeter.
3. Remove rubber holster.
4. Remove screw from battery
cover, battery cover, and battery.
5. Remove screws from back of
multimeter.
6. Remove back cover.
7. Remove fuse.
8. Replace fuse with same size
and type as originally installed.
Fuse 1: 500mA, 250V fast type
ceramic fuse Ø5 x 20mm.
Fuse 2: 10A, 250V, fast type,
ceramic fuse, Ø6.35 x 31.8mm.
9. Re-assemble multimeter.