4.2.2 Internal Laser warning information

Location of Laser warning
Positioned in the inside rear of
the device.
Positioned on the beam cover.
Positioned on the SLD cover

4.3 Electromagnetic compatibility (EMC)

The P200TE is intended for use in the electromagnetic environment as specified in the following tables. The
customer or the user of the device should assure that it is used in such an environment.
4.3.1 Guidance and manufacturer declaration - electromagnetic
emissions

TABLE 13: List of electromagnetic emission compliance

IEC60601
Emissions test
compliance
test level
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC61000-3-2
Voltage fluctuations/
Pst
flicker emissions
Dmax
IEC61000-3-3
4.3.2 Guidance and manufacturer declaration – electromagnetic
immunity

TABLE 14: List of electromagnetic immunity compliance

IEC60601 compliance test
Immunity test
level
Electrostatic
±6 kV contact
discharge
±8 kV air
IEC61000-4-2
Part Number: G108707/8GWE
Copyright 2018, Optos plc. All rights reserved.
Warning information
CAUTION - CLASS 3B LASER RADIATION
WHEN OPEN AVOID DIRECT EYE EXPOSURE
VISIBLE AND INVISIBLE LASER RADIATION
AVOID DIRECT EXPOSURE TO BEAM
CLASS 3B LASER PRODUCT
WARNING - LASER RADIATION
AVOID EXPOSURE TO BEAM
CLASS 3B LASER PRODUCT
Electromagnetic environment-guidance
The P200TE uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are
cause any interference in nearby electronic equipment.
The P200TE is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Electromagnetic environment-guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Chapter 4 - Technical Specification