Wöhler CD 210
6. Auto Power Off
The meter turns off automatically after 20 minutes of inactivity. To override
the function, hold down the ON/OFF-key and HOLD for two seconds to
turn the meter until "n" appears.
Note: Auto sleep function will be disabled during the calibration mode.
7. Setup
Hold down the ON/OFF-key under normal mode for more than one second
to enter the setup mode. To exit setup, press CAL/ESC in P1.0 and P3.0-
mode (see chapter 7.1) and it returns to normal mode.
7.1. CO
-Alarm (P1.0)
2
When entering setup mode, P1.0 and "Al" are shown in the display (Fig.
7.1). Press M/AV to go into P1.1 for setting the CO
-alarm threshold. The
2
current set value will be blinking on the display (Fig. 7.2).
Press MODE to increase the value or DP/WBT to decrease it. Each press
tunes 100 ppm and the alarm range is from 100 to 9900 ppm. When the
preferred alarm value is set, press M/AV to save the setting or CAL/ESC
without saving and return to P1.0
Fig. 7.1
Fig. 7.2
7.2. Temperature unit (P3.0)
Press MODE or DP/WBT in P1.0 to access P3.0 for setting up temperature
scale. Press M/AV and it goes into P3.1 with blinking °C or °F current set
on the lower left display. To switch °C or °F press MODE and DP/WBT.
Then press M/AV to save the settings or CAL/ESC to return to P3.0 with-
out saving.