5 Operation

5Operation
WARNING! Dangers will result from damage, unusual operating noises, excessively strong vibrations, excessive heating or if the
cutter or grinder are not tightly fastened!
Dangers to the user and patient.
 Discontinue use and inform the service department.
WARNING! Risk of burns from the hot head or lid of the instrument!
If the instrument overheats, burns may occur in the mouth area.
 Never touch soft tissue using the head or lid of the instrument.
CAUTION! Risk posed by an incorrectly stored instrument!
Clamped cutters or grinders may lead to injury or infection.
Damage to the chuck may occur if the instrument is dropped.
 Correctly return instrument to the tray without the cutter or grinder after the treatment.
NOTE! Premature wear and malfunctions may arise as a result of incorrect storage before long periods of non-use!
Shorter product life cycle.
 Before long periods of non-use, the medical product should be cleaned according to the instructions, treated and stored in a dry
location.