3 Reprocessing methods
Option I
Manual exterior cleaning
Necessary accessories:
• Potable water 30 ° C ± 5 ° C (86 ° F ± 9 ° F)
• Brush, e.g., medium-hard toothbrush
Brush under running potable water for
20 seconds.
Manual interior cleaning and maintenance (for U.S.
market)
Use STATCARE spray in accordance with the instructions for use.
Page 19
Option II
Automated internal and external cleaning and
disinfection in a washer-disinfector according
to ISO 15883
NOTE! Wet medical product after reprocessing!
Defects with the use of wet medical product.
 Choose suitable cleaning and drying method.
 Ensure that the medical product is completely dry after reprocessing.
In some countries the applicable national guidelines require
automated reprocessing in a validated washer-disinfector (WD)
such as, for instance, SciCan HYDRIM M2 or C61! Observe
instructions for use.
Validated parameters for thermal disinfection
Detergent:
PH-Value:
Program:
Temperature:
Exposure time:
Drying time:
TPH 5949 (S&M)
max. 10
Thermal disinfection
≥ 93 °C
≥ 5 minutes
≥ 15 minutes