3 Reprocessing methods
Recommended sterilization methods or parameters are as follows:
Pressure impulse method
S cycle
Vacuum method B cycle
* Depending on the type of use. National guidelines have to be observed.
Follow the instructions of the autoclave manufacturer regarding the sterilization of this medical products for autoclaves with S cycle method.
Pursuant to U.S. and Canadian standards:
Dynamic Air Removal:
Unwrapped
Dynamic Air Removal:
Wrapped or pouched
NOTE! This table represents the variation in sterilizer manufacturers' recommendations for exposure at different temperatures!
 For a specific sterilizer, consult only that manufacturer's recommendations.
Page 22
134 ° C (273.2 ° F)
Minimum pressure 300 kPa
134 ° C (273.2 ° F)
Minimum pressure 300 kPa
132 ° C (270 ° F)
Minimum sterilization time = 3 minutes
134 ° C (273 ° F)
Minimum sterilization time = 3.5 minutes
135 ° C (275 ° F)
Minimum sterilization time = 3 minutes
132 ° C (270 ° F)
Minimum sterilization time = 4 minutes
134 ° C (273 ° F)
Minimum sterilization time = 3.5 minutes
135 ° C (275 ° F)
Minimum sterilization time = 3 minutes
Minimum
3.5 - 5 - 18 minutes*
sterilization time
3.5 - 5 - 18 minutes*
Minimum
sterilization time
Minimum
15 minutes
drying time
10 minutes
Minimum
drying time
Drying time not applicable
Drying time not applicable
Drying time not applicable
Drying time = 60 minutes
Drying time = 60 minutes
Drying time = 60 minutes