Reprocessing Methods; Preparation At The Location Of Use; Cleaning - SciCan SANAO 200 L Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3 Reprocessing methods

3Reprocessing methods
Pos : 19 /926-Syc oT ec/3_Aufber eitungs methoden/2_Aufber eitungs methoden (H ands tüc ke und T urbine) @ 9\mod_1375798912385_6.doc @ 81645 @ 22333333232 @ 1
In the following the reprocessing methods according to ISO 17664 are described.

Preparation at the location of use

WARNING! Danger from non-sterile products!
Contaminated medical products pose a threat of infection.
 Take suitable action to protect personnel.
 Immediately remove residues of cement, composite or blood at the place of use.
 Transport the medical product in dry condition for reprocessing.
 Do not soak in solutions.
 The medical product must be reprocessed right after the treatment.

Cleaning

NOTE! Malfunctions by cleaning in ultrasonic device!
Defects of the medical product.
 The medical product must not be placed in a cleaning basin or ultrasonic device.
Page 18
SANAO
EN

Publicité

Table des Matières
loading