4 Installation and Commissioning
4Installation and Commissioning
WARNING! Risks posed by unsterilized products!
Risk of infection to user and patient.
Before putting the device into operation for the first time and after each use, reprocess the medical product accordingly and sterilize if
necessary.
NOTE! Malfunction of the handpiece as a result of a low ambient temperature!
Possible malfunctions of the handpiece.
Before putting into operation, return cooled products to a temperature of 20 – 25 ° C (68 – 77 ° F).
NOTE! Damage may occur as a result of polluted and moist cooling air!
Polluted and moist cooling air may lead to malfunctions and premature bearing wear.
Ensure that the cooling air is dry, clean and unpolluted as per ISO 7494-2 standards.
Cooling media supply
CAUTION! Overheating of tooth as a result of a low volume of water!
Thermal damage of the pulp.
Set water volume to at least 50 ml/min for the spray cooling.
Inspect spray water channels and clean spray nozzles if necessary.