4 Installation and Commissioning

4Installation and Commissioning
WARNING! Risks posed by unsterilized products!
Risk of infection to user and patient.
 Before putting the device into operation for the first time and after each use, reprocess the medical product accordingly and sterilize if
necessary.
NOTE! Malfunction of the handpiece as a result of a low ambient temperature!
Possible malfunctions of the handpiece.
 Before putting into operation, return cooled products to a temperature of 20 – 25 ° C (68 – 77 ° F).
NOTE! Damage may occur as a result of polluted and moist cooling air!
Polluted and moist cooling air may lead to malfunctions and premature bearing wear.
 Ensure that the cooling air is dry, clean and unpolluted as per ISO 7494-2 standards.

Cooling media supply

CAUTION! Overheating of tooth as a result of a low volume of water!
Thermal damage of the pulp.
 Set water volume to at least 50 ml/min for the spray cooling.
 Inspect spray water channels and clean spray nozzles if necessary.