Hach QP 1680-TOC Manuel D'utilisation De Base page 3

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22
Table of Contents
1 Specifications
on page 3
2 Online user manual
3 General information
4 Installation
on page 8
Section 1 Specifications
Specifications are subject to change without notice.
Specification
Dimensions (W x H x D)
Weight
Pollution degree
Overvoltage category
Parameters
Oxidation method
Temperature
Measuring range
Detection limit
Repeteability
Sample volume
Power requirements
Operating temperature
Storage temperature
Relative humidity
Altitude
Environmental conditions Indoor use
Gas supply
Gas connection
1
Supplied with the analyzer.
on page 4
on page 4
Details
38 x 44 x 70 mm (15.0 x 17.4 x 27.6 in.)
29 to 32 kg (64.0 to 70.6 lb)
2
II
TOC, TNb
TOC: Non-dispersive infrared detection (NDIR); TNb:
Chemiluminescence
Furnace temperature 1050 °C (1922 °F) maximum
0 to 30,000 mg/L
TC, TIC, NPOC: 50 µg/L; TNb: 20 µg/L
Until 10 mg/L: < 5%; More than 10 mg/L: < 2%
10 to 1000 µL
Analyzer: 100-240 VAC ±10%, 50/60 Hz; 16 A (with safety ground);
750 W maximum
PC: 100-240 VAC ±10%, 50/60 Hz; 16 A (with safety ground); 90 W
maximum
Monitor: 100-240 VAC ±10%, 50/60 Hz; 16 A (with safety ground);
100 W maximum
20 to 30 °C (68 to 86 °F)
-20 to 60 °C (-4 to 140 °F)
20 to 80%, non-condensing
2000 m (6562 ft) maximum
Type: Oxygen or synthetic air; Quality: 99,998% minimum (4.8);
Pressure: 3 to 10 bar (43.5 to 145 PSI)
1/8-inch Swagelok male connector for 1/8/-inch OD tubing
5 Startup
on page 15
6 Maintenance
on page 17
7 Troubleshooting
on page 19
1
English 3

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Qp 1680-tnb

Table des Matières