Electromagnetic Compatibility (EMC)

Electromagnetic emissions and test levels
The product is intended to be used in the stated environments with electromagnetic levels as specified below. The
caregiver and/or user of the product assure that the product is used in such an environment.
Emission test
RF emissions
RF emissions
Hamonic current emissions
Voltage fluctuations and Flicker
emissions
Electromagnetic Immunity
Acceptance criteria for the EMC pass criteria is unintentional movement above ±10mm is not allowed. The product
is intended to be used in the stated environments with electromagnetic levels as specified below. The caregiver
and/or user of the product assure that the product is used in such an environment.
Acceptance criteria for the EMC pass criteria is unintentional movement above ±10mm is not allowed. The product
is intended to be used in the stated environments with electromagnetic levels as specified below. The caregiver
and/or user of the product assure that the product is used in such an environment.
Radiated Fields in close proximity, Immunity test and compliance.
Standard: IEC 61000-4-39
Test levels (A/m)
8
65
7.5
136
Standards
CISPR 11
CISPR 11
IEC 61000-3-2
IEC 61000-3-3
Dwell time: 3 sec
Modulation
Pulse modulation
CW
50% Duty cycle
50% Duty cycle
Compliance
Group 1
Class B
Class A
Complies
Mod. Frequency (kHz)
CW
2.1
50
Test Frequency
30 kHz
134.2 kHz
13.56 MHz
I F U