Error Messages - Baxtran SML Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Validate using the key
0
It is going to appear the message "Load"
Hang the calibration mass and wait until the LED diode which indicates stability switches on.
Press the key
After having done the appropriate adjustment the self-testing is going to start. And next the weighing hook will automatically go back to
the weighing mode. If there is an adjustment error or a wrong mass of calibration, a message of error is going to appear on the screen
If you want to leave the calibration process at any moment, please press the key

8. ERROR MESSAGES

ERROR MESSAGE
-----
Err 4
Err 5
Err 6
Err 9
--oL--
Fail H
Fail L
bA Lo
Lo bA
If these messages appear in the screen, you must follow the recommendations, switch off and switch on the weighing hook. If the error message
does not disappear, please contact your distributor.
24
.
DESCRIPTION
The limit of load has been exceeded
Zero error
Keyboard error
Value of the A/D converter out of range
Unstable reading
Overload
Error of calibration
Discharged battery
USER'S MANUAL SML
.
RECOMMENDATIONS
Reduce the load.
Verify that the weighing hook has not suffered
any damages.
The deviation of zero exceeds the 4% of the
maximum capacity. Make sure that the hook is
unloaded.
Check the keys
Make sure that the hook is unloaded.
Check that there aren't strong wind currents,
vibrations, electromagnetic noise or something
is touching the hook
Remove the load
Recalibrate if this was inferior to the maxim
Recalibrate
Recharge the battery
EN
.

Publicité

Table des Matières
loading

Table des Matières